NOx/O2 / NH3 on-line detection system
AIREP is committed to the development and industrialization application of flue gas purification technology
Product features
1, High precision, high resolution, rapid response (the actual response time is within 2 seconds), good stability and repeatability;
2, Automatic tracking of zero and full range, and temperature compensation can be performed;
3, Automatic protection functions, including waterproof, explosion-proof, shock proof, and high concentration gas percussion, as well as good anti-jamming performance;
4, Software calibration function. Users can also calibrate by themselves. Simple operation and convenient maintenance;
5, Data recovery function, which can exempt you from worries caused by wrong operation;
6, High precision inner and outer filter, which can filter impurities effectively;
7, Adopt imported high-performance sensors, and the service life is up to 3-5 years;
8, Regular counter blowing function which can help to clean automatically.
Application range of NOx/O ₂ / NH ₃ detector
NOx/O₂/NH₃ detectors are widely used in the flue gas emission monitoring in electric power, glass, steel, chemical engineering,
kiln, environmental monitoring, distributed energy, research center, experimental platform and other fields.
The detector consists of NOx online detector and ammonia escape analyzer. It can perform real-time detection of NOx, and
convert according to the actual oxygen content. At the same time, it can also detect the escape amount of NH3 in the exhaust
gas. The detector adopts the extraction method, which separates the sampling system from the analysis system, so the analysis
system is in a space with relatively good environment.
The collected data can be converted into 4-20 mA output. It can be extended to RS485 output. It can also be extended to touch
screen. At the same time, it has internet of things remote transfer function, which can transmit the field monitoring data to all
over the world through GPRS/wifi. It is convenient for after-sales maintenance and monitoring the actual usage.
As the flue gas discharged by industrial enterprises often contains dust, particulate matter, ammonium sulfate, ammonium
hydrogen sulfate and other sulfide, which will cause blocking, short circuit or poisoning to the probes, the flue gas need to be
filtered. In order to prevent water vapor condensation and soot blocking, this equipment is equipped with heating belts and
ceramic filters to effectively filter impurities.
Good faith
innovation
AIREP is committed to the development and industrialization application of flue gas purification technology
Measuring range:NOx:0-3000ppm, O₂: 0-21%, NH₃: 0-50ppm
Measuring accuracy:≤±5%
Response time:≤2s
Environment temperature requirements:-20〜50 °C
Flue gas pressure:-6〜1kpa
Flue gas temperature:0〜800°C
Response time:2s
Sampling probe weight:about 30kg
Analysis tank weight:about 80kg
Signal output:4-20mA or 0-5V
Internet of things function:
It also has the GPRS output NOx, O ₂, NH ₃ analog quantity function. Relying on a
lifetime free open internet of things cloud platform, it can realize remote real-time
view data, record export, alarm information push. It can change the frequency, and
set physical value corresponding to the analog quantity in the Internet of things
platform. Real time curves and history records can be viewed on the mobile web
pages, WeChat or computer web pages.
Decomposition principle of nitrogen oxide